IBUS Guidelines |
Equipment
Examination technique
Interpretation
Linear array or annular array transducer configurations are suitable for high quality images.
The dominant transducer frequency should be 7.5
MHz or higher, preferably of a broad bandwidth construction .
However, high frequency on its own is not a sufficient
parameter to ensure quality.
A field of view greater than 4 cm is preferable for large area examinations, and smaller fields are suitable for detailed examination of specific findings.
A penetration depth of at least 4 cm with selectable focal regions is required.
The transducer dead zone should be minimal, and limited to distance less than 2-3 mm to clearly display superficial structures.
When colour Doppler imaging is included in the examination procedure, the colour display only provides an estimate of vascularity.
High sensitivity for low amplitude and low signal flow detection requires Doppler frequencies above 5 MHz.Low flow velocity detection requires wall filter settings in the range 50 - 100 Hz, and low pulse repetition frequency (PRF) around 1000 Hz.
Systematic
ComprehensiveReproducible
| systematic | A planned approach for performance and documentation
of the examination is required.
When a specific lesion is examined, its precise position should be noted on the image and its correlation with the clinical and mammographic findings recorded. When the total breast volume needs to be assessed, overlapping scans will ensure complete examination. |
| comprehensive | All breast structures must be completely displayed, and particular care taken in the area of interest to ensure that normal anatomical and/or pathological findings are recognized and recorded. |
| reproducible | The imaging results must be readily reproducible, and ultrasonic findings should be clearly identified on the stored images. It should be possible to confirm the same appearances on different types of high resolution ultrasonic scanning systems. |
As the acquisition of ultrasonic images is very operator dependent, a thorough understanding of the physical principles of ultrasound and of the normal anatomy are essential to achieve high quality images.
Supine oblique or supine position is recommended to reduce breast thickness and to improve visualization of deeper tissues. The reduced thickness allows optimization of focusing.
One or both arms should be elevated behind the head or neck to stretch the pectoralis muscle for better fixation and immobilization of the breast.
When scanning, the transducer should always be perpendicular to the skin surface.
Transducer coupling to the skin surface should be gentle and should give complete contact.
Strong compression pushes lesions out of the scanning plane below the transducer and should be avoided as it deforms tissue structures making interpretation more difficult.
Compression is useful to avoid refraction and scattering from normal anatomical structures when sound penetration is insufficient, and to examine tissue elasticity of benign and malignant findings.
The scanning procedure should involve overlapping scanning planes. These may be parasagittal, transverse, radial, or antiradial. Radial and antiradial scans follow normal anatomical patterns, assisting the recognition of abnormalities and better demonstrating ductal structure and changes.
a) Detailed knowledge of anatomy, physiological changes and benign and malignant pathology.b) Correlation of findings with other imaging results, clinical information and examination.
| Lesion features | Adjacent features |
| Shape (including depth/width ratio) | Architectural pattern |
| Boundaries/margins (including edge refraction) | Fibroglandular echogenicity |
| Internal echo texture | Fat echogenicity |
| Through transmission (enhancement, shadowing) | Cooper's ligament thickness |
| Calcifications | Skin thickness |
| Ductal alterations |
The indications for the examinationA description of any lesion(s) and adjacent features including the size of maximum diameter(s) or extent
The position of the lesion(s) as represented on a clock face with its distance from the nipple
Correlation with clinical and/or mammographic or other imaging findings
Opinion regarding provisional diagnosis(es) and significance of finding(s)
1) Cytology and core biopsyUltrasound guidance, either free hand or with mechanical guidance, is best performed with the needle shaft parallel with the transducer length so that the full length of the needle and the tip can be visualized at all times.2) Abscess drainage
3) Preoperative localization: hookwire placement, carbon tracking, and radionuclide marking
4) Intraoperative localization or localization
5) Radionuclide injection for sentinel node identification
6) Specimen imaging for verification of lesion removal
Free hand puncture enables the angle of approach to be optimized. Where practical, the angulation between the needle shaft and the transducer contact surface should be keep to a minimum. A horizontal angle of approach, with the needle positioned approximately parallel with the chest wall, is required for safety reasons and also for better angulation of the needle to the beam, improving the needle display.
Experience with interventional procedures is required for expertise.
Continuous education and follow up are essential to improve and maintain skills in technique and interpretation.
Recommendations by the International Breast Ultrasound School (IBUS) to achieve accuracy and confidence: